Abstract

To investigate MRI of the fetal heart by way of a novel triggering method with the use of an MR-compatible cardiotocography (CTG) in an animal model. Fetal cardiac MRI was performed on four pregnant ewes on a 1.5 Tesla (T) MR system. A CTG was rendered MR compatible and its signal was used for the triggering of the fetal heart to perform cardiac cine MRI of the fetal heart with maternal free-breathing with cine steady-state free precession. The left ventricular volume and function were measured from the short-axis (view). The image quality of anatomical structures was assessed. All cardiac valves and the foramen ovale could be visualized. Myocardial contraction was depicted in cine sequences. The average blood volume at the end systole was 1.7 mL (SD ± 0.12). The average volume at the end diastole was 4.6 mL (± 0.4); thus the average stroke volumes of the left ventricle were 2.87 mL (± 0.31) with ejection fractions of 60.53% (± 4.17). The newly developed MR compatible CTG could be used as a tool for cardiac triggering method of the fetal heart. This novel device might help fetal cardiac MRI technology in the future.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.