Abstract

A technique is described for high speed interactive imaging of the heart with either white or black blood contrast. Thirty-two views of a segmented, magnetization-prepared gradient echo sequence are acquired during diastole. Using three-quarter partial Fourier sampling, data for a complete 128 x 128 image are acquired in three cardiac cycles. High speed reconstruction provides an image update of each cardiac cycle 159 ms after measurement. An independent graphical user interface facilitates interactive control of section localization and contrast by permitting pulse sequence parameter modification during scanning. The efficiency and image quality of the cardiac MR fluoroscopy technique were evaluated in 11 subjects. Compared with the conventional graphic prescription method, the cardiac fluoroscopy technique provides an approximate eightfold reduction in the time required to obtain subject-specific double oblique sections. Image quality for these scout acquisitions performed during free breathing was sufficient to identify small cardiac structures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.