Abstract

Background and ObjectivesAlthough magnetic resonance imaging (MRI) conditional cardiac implantable electronic devices (CIEDs) have become recently available, non-MRI conditional devices and the presence of epicardial and abandoned leads remain a contraindication for MRIs.Subjects and MethodsThis was a single center retrospective study, evaluating the clinical outcomes and device parameter changes in patients with CIEDs who underwent an MRI from June 1992 to March 2015. Clinical and device related information was acquired by a thorough chart review.ResultsA total of 40 patients, 38 with a pacemaker (including epicardially located pacemaker leads) and 2 with implantable cardioverter defibrillators, underwent 50 MRI examinations. Among the patients, 11 had MRI conditional CIEDs, while the remaining had non-MRI conditional devices. Among these patients, 23 patients had traditional contraindications for an MRI: (1) nonfunctional leads (n=1, 2.5%), (2) epicardially located leads (n=9, 22.5%), (3) scanning area in proximity to a device (n=9, 22.5%), (4) devices implanted within 6 weeks (n=2, 5%), and (5) MRI field strength at 3.0 Tesla (n=6, 15%). All patients underwent a satisfactory MRI examination with no adverse events during or after the procedure. There were no significant changes in parameters or malfunctioning devices in any patients with CIEDs.ConclusionUnder careful monitoring, MRI is safe to perform on patients with non-MRI conditional CIEDs, remnant leads, and epicardially located leads, as well as MRI-conditional devices.