Abstract

The study investigated radiation dose, vascular computed tomography (CT) enhancement and image quality of cardiac computed tomography angiography (CCTA) with and without bolus tracking (BT) methods in infants with congenital heart disease (CHD). The volume CT dose index (CTDIvol) and dose length product (DLP) were recorded for all CT scans, and the effective dose was obtained using a conversion factors. The CT number for the ascending aorta (AO) and pulmonary artery (PA), image noise of muscle tissue and contrast-to-noise ratio (CNR) were measured and calculated. The median values in the groups with and without BT were 2.20mGy versus 0.44mGy for CTDIvol, 8.10mGy·cm versus 6.20mGy·cm for DLP, and 0.66mSv versus 0.51mSv for effective dose (p<0.001). There were no statistical differences in vascular CT enhancement, image noise, and CNR. CCTA without BT methods can reduce the radiation dose while maintaining vascular CT enhancement and image quality compared to CCTA with BT methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.