Abstract

Mechanical chest compression may be necessary to make coronary intervention possible during resuscitation. We report our experience using the Lund University Cardiac Arrest System (LUCAS, Jolife, Lund, Sweden) which is a gas-driven sternal compression device that incorporates a suction cup for active decompression. During the last 13 months LUCAS has been used in our catheterisation laboratory to maintain adequate organ perfusion pressure in 13 patients with cardiac arrest or severe hypotension and bradycardia (male/female ratio 1.6, mean age 59+/-19). The mean compression time was 105+/-60min (range 45-240), and the mean systolic and diastolic blood pressure obtained was 81+/-23 and 34+/-21mmHg, respectively. Angiography and eventually percutanous coronary intervention was possible in all cases during ongoing automatic chest compression. Three patients survived the procedure, but no patients were discharged alive. In two cases we found inadequate flow in the anterior descending artery, and in one case the invasive measurements revealed inadequate coronary perfusion pressure. There were no excessive intra-thoracic or intra-abdominal injuries. We conclude that the LUCAS device is suitable during cardiac catheterisation and intervention, and the device ensures an adequate systemic blood pressure in most patients without life-threatening injuries.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call