Abstract

The carbon response characteristics of a recently noted micro-flame ionization detector (μFID) mode are examined in detail. The μFID supports an extremely small (30 nL) “upside-down” flame that is generated from a low counter-current flow of oxygen immersed in hydrogen. Ionization measurements made in the μFID are directly compared to those obtained from a conventional FID. In terms of reproducibility of response and relative sensitivity towards different types of hydrocarbons, the μFID and a conventional FID produce no major differences with respect to either of these characteristics for a variety of compounds examined. Of note, for replicate measurements made in each detector, the average %R.S.D. of response typically differs by less than 2% between the two devices, while the average normalized sensitivity differs by less than 4%. In contrast to this, regarding absolute sensitivity, the analyte signal from the conventional air-rich FID is found to be three times larger than that of the hydrogen-rich μFID mode explored here. This discrepancy is ascribed directly to the difference in flame stoichiometry between the two detectors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.