Abstract
Part of the patella ligament was resected in rabbits and replaced by carbon fibre implants. After 1, 2, 3, 6 weeks, 3 and 12 months carbon implants with surrounding tissues as well as iliac lymph nodes were removed and examined by light- and transmission electron microscopy. Of special interest were the questions whether the carbon fibre implant, will be removed by phagocytosis and substituted by a new ligament or a adequate repair tissue? Our light- and transmission electron microscopic studies give no indication of successful removal of carbon fibres by phagocytosis. The carbon fibre implant was surrounded by dense connective tissue like a scar. In deep layers of the carbon implant was no vital dense or regular connective tissue, even not after 3 and 12 months. There is no replacement of the carbon fibre implant by a new ligament or tendon but a persistent foreign body reaction. With regard to these results it is very doubtful if good late results with ligament and tendon replacement by carbon fibre implants can be expected in patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.