Abstract

Electrospray ionization (ESI) is one of the most widely used ionization techniques, and its simplification can benefit many interested users. In this study, sample introduction by capillary action was studied and used to develop a simple ESI source called capillary self-aspirating electrospray ionization (CSESI). A conventional CSESI source requires only a common capillary of appropriate diameter in addition to the support of high voltage (HV). No pumps and sample loading is needed because the solution can spontaneously climb across the capillary by capillary action. With the proper modification of the glass capillary, the operation of CSESI can be further simplified and efficient. Specifically, cold plasma processing of the capillary creates a more hydrophilic surface that can facilitate sample introduction. Moreover, sputtering a thin platinum layer on the capillary tip makes the application of HV more convenient, and it also eliminates the influence of air bubbles in the capillary to ensure a sustained and stable electrospray. Overall, CSESI exhibits multiple desirable features such as simple structure, self-aspiration ability, low sample consumption, and inherent physical filtration capability. Apart from the routine ESI-MS analysis, it has also been applied in real-time monitoring of the oxidative dimerization of 8-methyl-1,2,3,4-tetrahydroquinoline, as well as direct analysis of muddy soil solutions without pretreatment.