Abstract

This work presents several critical details for making cIEF-MALDI-MS a robust technique which will allow for more routine application and aid in automation. This includes emphasis on the hardware necessary for syringe pump mobilization and proper protocol for preventing disruption from gas bubbles. Following these guidelines, excellent elution time reproducibility is demonstrated for six pI markers (RSD <5%). Additionally, the pI markers are used to calibrate the pH gradient and determine experimental pIs of proteins detected offline by mass spectrometry. This was demonstrated using a standard protein mixture of myoglobin and two forms of beta-lactoglobulin. Experimental determination of protein pIs and molecular weights were found to be in agreement with literature values. The technical details discussed provide a sound foundation for applying the offline coupling of MALDI-MS with cIEF.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.