Abstract
Capillary gas chromatography with atomic emission detection (GC-AED) was evaluated for the analysis of polychlorinated biphenyls (PCBs). Since Cl-responses were almost independent of the PCB structure, individual PCBs were quantitated with an accuracy better than 10% utilizing a Cl-calibration plot based on a single randomly selected congener (universal calibration). In addition, within a 5–10% accuracy, GC-AED enabled estimation of total PCB residue levels and calculation of the percentage by weight of chlorine in contaminating PCB mixtures. Thus, although PCB detection limits were higher with GC-AED than with GC-ECD, the former technique was very attractive for PCB investigations and enabled significant simplification of PCB quantitation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.