Abstract

A new method for rapid, direct determination of formate in blood serum samples by capillary electrophoresis with contactless conductometric detection is presented. A selective separation of formate was achieved in approximately 1min using an electrolyte system comprising 10mM l-histidine, 15mM glutamic acid and 30μM cetyltrimethylammonium bromide at pH 4.56. The only sample preparation was dilution (1:100) with deionized water. The limit of detection and limit of quantitation was 2.2μM and 7.3μM, respectively, which corresponds to 0.22mM and 0.73mM in undiluted blood serum. The method provides a simple and rapid diagnostic test in suspected methanol intoxication cases. The method has been successfully tested on determination of formate in blood of a patient admitted to the hospital under acute methanol intoxication. The peak concentration of formate detected in the patient blood serum was 12.4mM, which is 10- to 100-fold higher than the normal values in healthy population. The developed method presents the fastest test currently available to detect formate in blood samples.