Abstract

Ucon-coated columns have been used to isolate nucleotides, such as ribonucleotides, deoxyribonucleotides, and pyridine nucleotides, by capillary zone electrophoresis. This neutral, hydrophilic column coating (Ucon) significantly reduced macromolecule adsorption and electroosmotic flow, which provides the maximum resolution of the nucleotides separated under moderate buffer conditions (pH about 5-6), The relative standard deviations of the nucleotides were less than 1%. Low minimum detectable concentrations (about 2-8 μM) and quantities (about 70-360 fmol) of the ribonucleotides were obtained. A qualitative and quantitative analysis of the ribonucleotides in hybridoma cell extracts shows the applicability of the developed methodology for the determination of intracellular nucleotide pools under the desirable buffer conditions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.