Abstract

It is commonly thought that even a moderately high ionic concentration in the background electrolyte (BGE) would lead to Joule heating and serious peak distortion. However, we obtained very satisfactory separations of both inorganic and organic anions in electrolyte solutions as high as 5 M sodium chloride using direct photometric detection. Samples containing a 0.5 M concentration of a salt can be analyzed directly by making the BGE concentration of the same salt even higher to obtain electrostacking. The temperature in the center of the capillary was calculated to be 49 degrees C when the current is at its maximum of 280 microA. The effect of various salts on electrophoretic and electroosmotic mobility is discussed. Several examples are given of capillary electrophoresis under high-salt conditions.