Abstract
The present work describes a method for the simultaneous determination of unmodified nucleosides and nucleotide mono-, di- and tri-phosphates by capillary electrophoresis coupled to mass spectrometry (CE-MS). The use of hexafluoro-2-propanol (HFIP) in the separation medium, and as an additive to the sheath liquid of the electrospray interface (ESI), generated a highly efficient and sensitive method. Instrumental limits of detection in the range of 14–53ngmL−1 for nucleosides and 7–23, 20–49 and 64–124ngmL−1 for nucleotide mono-, di-, and tri-phosphates, respectively, were found. Sample treatment involved diluting an aliquot of baby food with ultra-high quality water and applying centrifugation-assisted ultrafiltration (CUF). The proposed method was validated and used to analyse a variety of baby food samples (16 in total) such as fish, meat, fruits, and baby dairy desserts that may endogenously contain these analytes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.