Abstract

The time required for method development in gradient-elution liquid chromatography (LC) may be reduced by using an empirical modelling approach to describe and predict analyte retention and peak width. However, prediction accuracy is impaired by system-induced gradient deformation, which can be especially prominent for steep gradients. As the deformation is unique to each LC instrument, it needs to be corrected for if retention modelling for optimization and method transfer is to become generally applicable. Such a correction requires knowledge of the actual gradient profile. The latter has been measured using capacitively coupled “contactless” conductivity detection (C4D), featuring a low detection volume (approximately 0.05 μL) and compatibility with very high pressures (80 MPa or more). Several different solvent gradients, from water to acetonitrile, water to methanol, and acetonitrile to tetrahydrofuran, could be measured directly without the addition of a tracer component to the mobile phase, exemplifying the universal nature of the approach. Gradient profiles were found to be unique for each solvent combination, flowrate, and gradient duration. The profiles could be described by convoluting the programmed gradient with a weighted sum of two distribution functions. Knowledge of the exact profiles was used to improve the inter-system transferability of retention models for toluene, anthracene, phenol, emodin, sudan-I and several polystyrene standards.