Abstract

Liquid holdup measurement of cryogenic fluids is an area of considerable significance because of its inevitable occurrence in LNG transportation, rocket propellant delivery and superconducting equipment cooling, etc. To measure the liquid holdup of cryogenic two-phase flow, a capacitance sensor was carefully designed, which consists of a pair of optimized concave-electrode form with the electric circuit for the small capacitance detection. Four flow patterns were realized to evaluate the performance of the sensor in visualization experiments with liquid nitrogen and vaporous nitrogen. An image method was employed to calibrate the capacitance sensor, which led to a mathematical relationship between the capacitance and the liquid holdup. The results indicated that the obtained correlation between liquid holdup and capacitance satisfactorily coincided with the measured data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.