Abstract

Chemical ionization Orbitrap mass spectrometry (CI-Orbitrap) represents a promising new technique for gas-phase analysis in analytical and atmospheric chemistry mainly due to its very high mass resolving power. In this work, we performed the first side-by-side comparison between a CI-Orbitrap and the widely used atmospheric pressure interface time-of-flight mass spectrometry (CI-APi-TOF) using two different chemical ionization methods, i.e., acetate-ion-based (CH3COO-) and aminium-ion-based (n-C3H7NH3+) schemes. The capability of the CI-Orbitrap at accurately measuring low concentrations of gaseous species formed from the oxidation of α-pinene was explored. Although this study reveals a lack of linearity of the CI-Orbitrap when measuring product ions at very low concentrations (<1 × 106 molecules cm-3), very good agreement between both techniques can be achieved by applying a newly developed linearity correction. It is experimentally shown that the correction function is independent of the reagent ion used. Thus, accurate quantification of organic compounds at concentrations as low as 1 × 105 molecules cm-3 by the CI-Orbitrap can be achieved. Finally, by means of tandem mass spectrometry, the unique capability of the Orbitrap allows the direct determination of the binding energy of cluster ions between analyte and reagent ions, that is needed for the assessment of a chosen ionization scheme.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.