Abstract
A new National Electrostatic Corporation (NEC) 5MV accelerator mass spectrometer became operational at the Scottish Universities Environmental Research Centre (SUERC) in July 2002. It has 2 Cs sputter negative ion sources: a 134-sample source (S1) for the routine measurement of all species, and a hybrid source (S2) with 40 spaces for radiocarbon measurements with either graphite or CO2 samples. A number of performance tests on graphite samples have been carried out on both sources. A precison of better than 0.3% is feasible for modern samples on a routine basis. The 14C background of the machine and the graphite preparation process blank are 0.04 ± 0.01 and 0.16 ± 0.05 pMC, respectively, indicating that 14C dating back to ~50 kyr BP is possible. The normalized 14C values for a series of reference materials agree well with the IAEA, TIRI, and FIRI consensus values. Routine measurement of 14C has been underway since May 2003. Preliminary results of performance tests on the CO2 gas ion source are also reported.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.