Abstract

A compact and sensitive photoacoustic setup has been developed based on a recently demonstrated cantilever technique. A micromechanical cantilever transducer is attached to a cylindrical photoacoustic cell and the cantilever’s deflection is monitored with a compact Michelson interferometer. A commercial 1-Watt optical fiber amplifier was used to enhance the performance of the system. A normalized sensitivity of 1.4×10-10 cm-1 W Hz-1/2 was achieved in the detection of carbon dioxide at 1572 nm wavelength. Using 34 mW optical power from a DFB diode laser, the noise-equivalent detection limit for carbon dioxide at this wavelength is 4.0 ppm. Employing the fiber amplifier, we improved the sensitivity to yield measurement of sub-ppm concentrations.