Abstract

The objective of this study was to assess the 12-month clinical outcomes in patients with drug-eluting stent in-stent restenosis (DES-ISR) who were either predilated with non-compliant balloons (NCBA) and with additional scoring balloons (NCBA + SBA) prior to drug coated balloon (DCB) angioplasty. This monocentric, retrospective study included patients with DES-ISR who were routinely treated over a 2-year time span. Patients with stable angina and documented ischemia or selected forms of unstable angina due to a culprit DES-ISR lesion were analyzed. The primary endpoint was the clinically driven target-lesion revascularization (TLR) rate at 12 months. Secondary endpoints included postinterventional lumen gain and late lumen loss (LLL) at 6 months. The 12-month TLR rates in 124 patients who underwent either NCBA + SBA or NCBA only group were not different (17.3%, 9/52 vs. 11.6%, 8/69, P = 0.371) and low as compared to other comparable studies. The use of SBA led to equally high post minimal lumen diameters (MLD) in both treatment arms (NCBA 2.21 ± 0.33 vs. NCBA + SBA 2.18 ± 0.41, P = 0.868). We did not find a significant difference in late lumen loss (LLL) between both groups (0.50 ± 0.62 mm vs. 0.40 ± 0.46 mm, P = 0.468). Scoring Balloon Angioplasty can safely and effectively prepare DES-ISR lesions to render them suitable for DCB angioplasty with acceptable TLR and MACE rates.