Abstract

ObjectiveTriggered electromyography (t-EMG) for pedicle screw placement was introduced to prevent the misplacement of screws; however, its diagnostic value is still debated. This study aimed to clarify the diagnostic value of t-EMG and to compare thresholds. MethodsWe searched MEDLINE, EMBASE, and the Cochrane Library, and 179 studies were identified. Among them, 11 studies were finally enrolled. The pooled sensitivity, specificity, diagnostic odds ratio (DOR), and summary receiver operating characteristics (SROC) plots were analyzed. ResultsThe enrolled studies included 13,948 lumbar and 2070 thoracic screws. The overall summary sensitivity/specificity/DOR values of t-EMG were 0.55/0.97/42.16 in the lumbar spine and 0.41/0.95/14.52 in the thoracic spine, respectively, indicating a weak diagnostic value. However, subgroup analysis by each threshold value showed that the cutoff value of 8mA in the lumbar spine indicated high sensitivity (0.82), specificity (0.97), and DOR (147.95), thereby showing high diagnostic accuracy of identifying misplaced screws. ConclusionThe most useful application of t-EMG may be as a warning tool for lumbar pedicle screw malpositioning in the presence of positive stimulation at a threshold of ⩽8mA. Significancet-EMG by screw stimulation may be valuable in the lumbar region at a threshold of ⩽8mA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.