Abstract

A total artificial heart (TAH) goat survived for 360 days on the new automatic control method (1/R control), in which the cardiac output of the TAH can be controlled through the cardiovascular center by making it function by reflecting the beta-adrenergic reaction in peripheral vascular resistance. This is thought to be the only long-term, real-time, measurable parameter by which information on the activity of the cardiovascular center can be received directly by the TAH system. In this goat, the hemodynamic parameters (RAP, AoP, and so forth) were kept within physiologic limits when control was stable, and the cardiac output was automatically increased in response to exercise, not unlike that in the natural heart. There were no abnormal blood chemical or hormone data except at end stage. Based on these results 1/R can be considered a physiologic control method for a TAH.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.