Abstract

We aimed to evaluate the quality of various 3D T1-weighted images (T1WIs) of the stent lumen using a carotid stent phantom and determine the suitable T1WI sequence for visualization of the stent lumen after carotid artery stenting. The carotid stent phantom consisted of polypropylene tubes that mimicked common carotid arteries with and without stenting. On 1.5T and 3.0T MRI scanners, transverse T1WIs of the carotid stent phantom were obtained using 3D turbo spin-echo (TSE), 3D fast field-echo (3D-FFE), and 3D turbo field echo volumetric interpolated breath-hold examination (VIBE) under clinical conditions. The signal intensity ratio (SIR) was determined using the mean signal intensity of the stent lumen (SIstent) divided by the lumen without a stent in each T1WI. The SNR of the stent lumen (SNRstent) was calculated from SIstent divided by the standard deviation of the uniform region near the stent lumen. The 3D-FFE and VIBE had higher SNRstent than other T1WIs and clearly visualized the stent lumen. The 3D-TSE had the lowest SIR and SNRstent, preventing stent lumen visualization. T1WIs obtained using 3D-FFE and VIBE allows stent lumen visualization.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.