Abstract

Individual calibration (IC) for caries detection methods based on fluorescence is time-consuming, especially for paediatric dentists, if the calibration has to be performed tooth-by-tooth. However, it is not clear how this calibration actually interfere in laser fluorescence (LF) readings. This in vivo study was to verify the influence of different modes of IC on laser fluorescence (LF) readings. Ninety six occlusal and 95 buccal surfaces of 1st permanent molars were examined using LF device after IC performed on control (no IC), the examined teeth, a permanent incisor, a 1st primary molar or a 2nd primary molar. All modes of IC were performed in the same child. Wilcoxon test and Bland-Altman analysis were used to compare the readings. Intraclass correlation coefficients (ICC) were calculated. Laser fluorescence readings without prior calibration were higher than readings performed after any mode of IC and resulted in different values of ICC. After other IC modes, the LF readings were statistically similar. The absence of IC influences LF readings and LF reproducibility, but different IC methods can be considered in clinical practice.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.