Abstract

IntroductionA study was conducted to determine whether the anode heel effect can be used to influence optimisation of radiation dose and image quality (IQ) for AP pelvis radiography. MethodsATOM dosimetry phantom and an anthropomorphic phantom were positioned for AP pelvis. Using a CR system, images were acquired and doses were measured with phantom feet toward anode and then feet toward cathode. Exposure factors (kVp, mAs and SID) were systematically generated using a factorial design. Images were scored visually for quality using relative visual grading together with a 3 point Likert scale. Signal to noise ratio was also calculated as a physical measure of image quality. Dosimetry data were collected for the ovaries and testes. ResultsThe optimum technique for male, which resulted in lower dose and suitable image quality, was with feet positioned toward the anode (0.80 ± 0.03 mGy; SNR of 38 ± 2.9; visual IQ score 3.13 ± 0.35). The optimum technique for female was with feet toward anode (0.23 ± 0.02 mGy; SNR of 34.7 ± 2.6; visual IQ score 3.15 ± 0.26). kVp had the biggest effect on both visual and physical image quality metrics (p < 0.001) for both tube orientations, whereas SID had the lowest effect on both visual and physical image quality metrics compared with mAs and kVp (p < 0.001). The effect of SID on the SNR was not significant (p > 0.05) with feet toward anode. ConclusionPositioning the patient with feet toward the anode, as opposed to the cathode, has no adverse effect on visual image quality assessment but it does have an effect on physical image quality. Implications for practiceThis study would add a new clinical concept in positioning of AP pelvis radiography especially for male positioning.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call