Abstract

Anecdotally, surgeons sometimes observe large errors when using image guidance in endonasal surgery. We hypothesize that one contributing factor is the possibility that operating room personnel might accidentally bump the optically tracked rigid body attached to the patient after registration has been performed. In this paper we explore the registration error at the skull base that can be induced by simulated bumping of the rigid body, and find that large errors can occur when simulated bumps are applied to the rigid body. To address this, we propose a new fixation method for the rigid body based on granular jamming (i.e. using particles like ground coffee). Our results show that our granular jamming fixation prototype reduces registration error by 28%-68% (depending on bump direction) in comparison to a standard Brainlab reference headband.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.