Abstract

According to new proposals from the American Diabetes Association (ADA) and WHO, venous peripheral plasma is the preferred system for measuring glucose for diagnosing diabetes mellitus. Owing to the instability of glucose in plasma after blood sampling, strict well-defined and standardized preanalytical conditions are essential to ensure that glucose concentration measured in plasma reflects real blood glucose in the patient. This is in contrast to the capillary whole blood measurements, which are easy to perform and well established. We investigated whether it is possible to perform analysis on capillary whole blood but express the results as plasma glucose values and hence obtain comparable results and the same predictive values for diagnosis in the individual patient? The conclusion of our investigations is that these two systems are not interchangeable and that conversion should not be done for diagnostic purposes where plasma determinations are recommended.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.