Abstract

The measurement of total body water (TBW)) by bioelectrical impedance (BEI) in a group of renal patients was evaluated against the tritium dilution method. The effect of haemodialysis and the presence of peritoneal dialysate on the impedance were also investigated. The correlation between the two methods was r = 0.9, with the standard deviation of the differences being 3.66 l (TBW typically 40 l). The BEI method overestimated the actual weight loss after haemodialysis, but underestimated the volume of peritoneal dialysate in situ. The BEI method would not be appropriate for use in assessing total body water and monitoring acute volume changes in patients with renal failure who are on strict fluid restriction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.