Abstract

BackgroundOne factor in the long-term survivorship of unicompartmental knee arthroplasty is the accuracy of implantation. In addition to implant designs, the instrumentation has also evolved in the last three decades to improve the reproducibility of implant placement. There have been limited studies comparing mobile bearing unicompartmental knee arthroplasty with contemporary instrumentation and fixed bearing unicompartmental knee arthroplasty with conventional instrumentation. This study aims to determine whether the Microplasty instrumentation in Oxford unicompartmental knee arthroplasty allows the surgeon to implant the components more precisely and accurately.MethodsA total of 150 patients (194 knees) were included between April 2013 and June 2019. Coronal and sagittal alignment of the tibial and femoral components was measured on postoperative radiographs. Component axial rotational alignment was measured on postoperative computer tomography. The knee rotation angle was the difference between the femoral and tibial axial rotation. A rotational mismatch was defined as a knee rotation angle of > 10°. Statistical analysis was performed using Student t test and Mann-Whitney nonparametric test. A p value < 0.05 was considered statistically significant in each analysis.ResultsBetween April 2013 to June 2019, 112 patients (150 knees) received Oxford unicompartmental knee arthroplasty, one patient (2 knees) had Journey unicompartmental knee arthroplasty, and 37 patients (42 knees) received Zimmer unicompartmental knee arthroplasty. All femoral components in the Oxford group were implanted within the reference range, compared with 36.6% in the fixed bearing group (p < 0.001). 88.3% of Oxford knees had tibial component falling within the reference range, whereas 56.1% of knees in the fixed bearing group fell within the reference range (p < 0.001). 97.5% of Oxford knees had tibial slope that fell within reference range, whereas 53.7% fell within range for fixed bearing group (p < 0.001). Femorotibial rotational mismatch of more than 10° was noted in 13.8% in Oxford group and 20.5% in fixed bearing group (p = 0.04).ConclusionIn conclusion, Microplasty instrumentation for Oxford mobile bearing unicompartmental knee arthroplasty is more accurate and precise compared to conventional fixed bearing unicompartmental knee arthroplasty in sagittal, coronal, and axial alignment. Prospective studies with long-term follow-up are warranted to investigate the clinical implications.

Highlights

  • Unicompartmental knee arthroplasty (UKA) is an established treatment option for patients with isolated medial or lateral compartment knee osteoarthritis

  • In contrast to the separate femoral and tibial saw guides in the conventional instrumentation in fixed bearing UKA, the phase III Oxford UKA Microplasty® instrumentation features a linkage of the tibial saw guide, the G clamp and the femoral sizing spoon; which reflects the axis of the medial femoral condyle, proposedly allowing for more consistent and accurate tibial resection [11]

  • Of the 194 medial UKAs performed between April 2013 and June 2019, 150 were mobile bearing and 44 were conventional fixed bearing

Read more

Summary

Introduction

Unicompartmental knee arthroplasty (UKA) is an established treatment option for patients with isolated medial or lateral compartment knee osteoarthritis. The Oxford mobile bearing UKA has made some changes to its instrumentation in an attempt to improve its accuracy and precision in implant positioning. In addition to higher instrumentation precision, the Oxford UKA has the theoretical advantage of adjusting for small amount of rotational mismatch, accepting a larger margin of error in implant positioning. One factor in the long-term survivorship of unicompartmental knee arthroplasty is the accuracy of implantation. There have been limited studies comparing mobile bearing unicompartmental knee arthroplasty with contemporary instrumentation and fixed bearing unicompartmental knee arthroplasty with conventional instrumentation. This study aims to determine whether the Microplasty instrumentation in Oxford unicompartmental knee arthroplasty allows the surgeon to implant the components more precisely and accurately

Objectives
Methods
Results
Discussion
Conclusion