Abstract
We present a new calorimeter for measuring the isobaric specific heat capacity for fluids and fluid mixtures based on the thermal relaxation method. The calorimeter, which is equipped with a metal bellows, works in the temperature range from 250to500K at pressures up to 20MPa and is designed to satisfy the conditions of a lumped system. The calorimeter was calibrated using water and toluene. The isobaric specific heat capacity of methanol was measured at 323.15K and 0.1MPa, which confirms the reliability of the calorimeter. The uncertainty is estimated to be within 2.9% at a 95% confidence level.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.