Abstract

The push broom design of an underwater hyperspectral imaging (UHI) instrument makes it possible to measure angle-resolved spectral radiance L(λ) in a plane. We describe the characterization of a commercial UHI instrument (UHI-4, Ecotone AS, Norway) and the spectral, geometric, and radiometric calibration transfer for measuring L(λ) in absolute units [µWcm−2nm−1sr−1]. We present a low-cost instrument characterization approach that is intended to be easily replicated for other users to perform their own calibration transfer. Cross-calibration with a RAMSES-ARC spectroradiometer (TriOS, Germany) in air and in water shows good linear correlation across the observed spectral range and over four orders of magnitude in dynamic range, with the UHI instrument providing higher sensitivity overall.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.