Abstract
We report on our efforts to provide high accuracy wavelength calibration to the scientific observations with the Cryogenic High-Resolution IR Echelle Spectrograph (CRIRES), ESO’s new high resolution (R≈100,000) infrared (IR) spectrograph at the VLT. In order to provide reliable and accurate wavelength standards for CRIRES the European Southern Observatory (ESO), in collaboration with the Space Telescope European Co-ordinating Facility (ST-ECF) and the National Institute of Standards and Technology (NIST), embarked on a project to establish Th–Ar wavelength standards in the 950nm to 5000nm operating range of CRIRES. Based on current findings we conclude that Th-Ar hollow cathode lamps hold the promise of becoming a standard source for wavelength calibration for near IR astronomy, providing a high density of sharp well-characterized emission lines with the ease and efficiency of operation of a commercial discharge lamp. In addition, and for use at wavelengths larger than 2200, we have established a gas cell filled with N2O as a calibration source based on existing data from NIST. Both sources were extensively tested during CRIRES commissioning runs and both are now being used for routine operations of the instrument.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.