Abstract

Calibration of ultrasonic hydrophone probes in the frequency range from 250 kHz to 1 MHz is required to sufficiently assess the peak rarefactional pressure (pr) and the Mechanical Index (MI) of medical ultrasound imaging devices. However, the ultrasonic hydrophone calibration in this low frequency is barely conducted. Therefore, the objective of this research was to develop a calibration technique for ultrasonic hydrophone probes in the frequency range from 250 kHz to 1 MHz. Two ultrasonic hydrophone probes, one membrane hydrophone and one needle hydrophone, were calibrated using a substitution method combined with time-delay spectrometry (TDS). The calibration results are presented in term of end-of-cable voltage sensitivity as a function of frequency. The calibration data show that the membrane hydrophone exhibit a very flat frequency response, to within ±1 dB for the entire investigated frequency range, whereas the needle hydrophone demonstrates a relative large variations in sensitivity of about 5 dB. These results are in good agreement with the limited data previously reported. Therefore, the substitution calibration technique with Time Delay Spectrometry (TDS) is capable of calibrating the ultrasonic hydrophone probes in the frequency range from 250 kHz to 1 MHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.