Abstract

This paper presents the ultrasonic flaw detectors calibration system at Standards and Calibration Laboratory (SCL) developed in accordance with the Group 2 tests of ISO 22232-1:2020. The calibration system contains a delay-pulse generator, a function generator, an amplifier, an oscilloscope and a set of tunable attenuators. The measurement setup, model and uncertainty components for each tests are presented. The difference in test parameters from the preceding international standard EN 12668-1:2010 are also reviewed and compared.