Abstract
Abstract We have constructed a beamline which serves for calibrating transfer standard light sources in the UV and VUV regions. The synchrotron radiation (SR) from the electron storage ring TERAS (750 MeV) is used as a primary standard of spectral radiant-intensity. In the beamline, a deuterium lamp was used for transfer standard light source. The SR is linearly polarized and the deuterium lamp is almost unpolarized, and hence the polarization effect of the optical system in its beamline cannot be neglected. We evaluated the polarization effect of the optical system and calibrated the spectral radiance of a deuterium lamp in the wavelength range 160–310 nm. The uncertainty of the spectral radiance of the calibrated lamp by the polarization effects is estimated to be about ±10%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Electron Spectroscopy and Related Phenomena
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.