Abstract

A calibration technique for a time domain reflectometry and transmission (TDR and TDT) measurement system as applied to network analysis is presented. The calibration corrects for the errors caused by the response of the measurement system. A complete physically-based model is established for the system. A set of calculable standards is developed to satisfy the time domain requirements for calibration. The calibration technique was applied to determining the model parameters of a commercial TDR and TDT system. The calibration enhanced the system bandwidth from 8 GHz to about 20 GHz. Experimental verification is given to demonstrate the validity and accuracy of the calibration technique. >

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.