Abstract

Pure beta particle emitting radionuclides have many therapeutic applications in nuclear medicine. The response of the NPL secondary standard radionuclide calibrator to 32P, 89Sr and 90Y has been measured using accurately calibrated solutions. For this purpose, high efficiency solid sources were prepared gravimetrically from dilute solutions of each radionuclide and assayed in a 4π proportional counter; the source activities were determined using known detection efficiency factors. Measurements were made of the current response (pA/MBq) of the NPL secondary standard radionuclide calibrator using the original concentrated solutions. Calibration figures have been derived for 2 and 5 ml British Standard glass ampoules and Amersham International plc P6 vials. Volume correction factors have also been determined. Gamma-ray emitting contaminants can have a disproportionate effect on the calibrator response and particular attention has been paid to this.