Abstract
The tagging system of the KEDR detector for the two-photon studies is briefly described. The procedure of the energy calibration of the system is considered. Two-photon events e+e− → e+e− +μ+μ− selected with the central detector were employed to check the energy calibration and to measure the tagging efficiencies. The results obtained demonstrate that the achieved accuracy of the energy calibration is about 1%. To improve the accuracy to the level of 0.3% the upgrade of the calibration system is ongoing. The tagging efficiencies obtained with the simulation agree with the measured one within the error of the measurement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.