Abstract

The results of cosmic muon calibration of the extended quadrihedral NaI(Tl) spectrometer are presented. The signal amplitude and energy resolution of the spectrometer in the direction transverse to the spectrometer axis have complex dependences on the photomultiplier voltage and particle entrance point to the spectrometer. At a photomultiplier divider voltage U = 1150 ± 10 V, the region of uniform signals is observed over ~90% of the spectrometer length, the relative energy resolution is δ ≈ 7%.