Abstract

Two methods were tested for calibration procedures of the 3-D coil system which is used as a source part of the magnetic tracker. Precise magnetometer's positions with respect to the coil system were used for determining of the nonorthogonality of the coil system as well as for determining of the magnetic moment. The needed number of the reference positions was measured to calculate six unknown calibrated parameters. In comparison, the 3-D coils system was also calibrated with a method which employs the orthogonal homogenous ac field generated in Helmholtz coils. The misalignment of up to two degrees was discovered.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.