Abstract

Four techniques for measuring the dose rate from Sr-90 concave eye plaques are presented. The techniques involve calibrating a concave eye plaque against a Sr-90 teletherapy unit using X-Omat film, radiochromic film, black LiF TLD discs and LiF chips. The mean dose rate predicted by these dosimeters is 7.5 cGy s-1. The dose rate quoted by the manufacturer is 33% lower than this value, which is consistent with discrepancies reported by other authors. Calibration against a 6 MV linear accelerator was also carried out using each of the above dosimetric devices, and appropriate sensitivity correction factors have been presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.