Abstract

Four techniques for measuring the dose rate from Sr-90 concave eye plaques are presented. The techniques involve calibrating a concave eye plaque against a Sr-90 teletherapy unit using X-Omat film, radiochromic film, black LiF TLD discs and LiF chips. The mean dose rate predicted by these dosimeters is 7.5 cGy s-1. The dose rate quoted by the manufacturer is 33% lower than this value, which is consistent with discrepancies reported by other authors. Calibration against a 6 MV linear accelerator was also carried out using each of the above dosimetric devices, and appropriate sensitivity correction factors have been presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.