Abstract
The performance of Wright peak flow meters is described using an absolute and precise calibration method. This showed errors as great as 25% with the peak flow meter and 40% with the mini peak flow meter. At large flow rates errors were less. The expected variation between individual peak flow meters was about 12%. The expected variation between individual mini peak flow meters was between 16% and 22%, depending on flow rate.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.