Abstract

In order to ascertain the spatial and temporal changes in the air quality in Temuco City, Chile, we created and installed a network of inexpensive sensors to detect PM2.5 particulate matter. The 21 measurement points deployed were based on a low-cost Sensiron SPS30 sensor, complemented with temperature and humidity sensors, an Esp32 microcontroller card with LoRa and WiFi wireless communication interface, and a solar charging unit. The units were calibrated using an airtight combustion chamber with a Grimm 11-E as a reference unit. The calibration procedure fits the parameters of a calibration model to map the raw low-cost particle-material measurements into reliable calibrated values. The measurements showed that the concentrations of fine particulate material recorded in Temuco present a high temporal and spatial variability. In critical contamination episodes, pollution reaches values as high as 354 µg/m3, and at the same time, it reaches 50 µg/m3 in other parts of the city. The contamination episodes show a similar trend around the city, and the peaks are in the time interval from 07:00 PM to 1:00 AM. In the winter, this time of day coincides with when families are usually home and there are low temperatures outside.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call