Abstract

The aim of this paper is to discuss the measurements carried out to implement ‘in vivo dosimetry’ with EDP-30 diodes in total body irradiation (TBI) techniques. Exit calibrations and calibrations behind cerrobend protection blocks showed the importance of calibrating diodes in all relevant clinical conditions. Special attention was given to calibration of diodes behind cerrobend blocks. Dependence of the calibration factors on the thickness of the shielding blocks was, therefore, studied. This dependence was again studied after adding a wax cap to the diode and when the ionisation chamber was placed at the same depth as the measuring point of the diode. Temperature dependence in diode sensitivity and dependence on accumulated dose for diodes response and for temperature correction factors were also examined.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.