Abstract

An intercomparison of field strength meters was performed during 1991-1993 over the frequency range 100 kHz-1000 MHz and field strength range 1-10 V/m using different calibration techniques. Four field sensors, electric as well as magnetic, have been used. Seven institutes participated in this intercomparison using at least three different calibration methods. A calibration uncertainty of less than 1 dB can be obtained if certain procedures are followed. The results are applicable in the fields of EMC testing and health care (nonionizing radiation). >

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.