The distortions introduced into high-resolution transmission electron microscope (HRTEM) images by the projector lens system are an important source of systematic error for quantitative displacement and strain determination. Using geometric phase analysis of images of perfect crystals, we measured these errors for two different transmission electron microscopes. Local magnification varies by as much as 5%, and rotation can reach 2 degrees across a typical image. Our experimental results are compared with theory, and optical pincushion and spiral distortion coefficients are determined. A method for calibrating and removing these distortions is presented that enables quantification to 0.1% strain and 0.1 degrees rotation across the whole field of view. This calibration is also critical for the accurate measurement of local lattice parameters from HRTEM images.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.