Abstract

The sampling of the surface-near gas composition using a mass spectrometer (MS-Probe) is a valuable tool within a hot dip process simulator. Since reference samples with well characterized surface coverage are usually not available, steel samples can deliver quantifiable amounts of the process relevant species H2O, CO and H2 using the decarburization reaction with water vapor. Such “artificial calibration samples” (ACS) can be used for the calibration of the MS-Probe measurements. The carbon release rate, which is governed by the diffusion law, was determined by GDOES, since the diffusion coefficients of carbon in steel samples are usually not known. The measured carbon concentration profiles in the ACS after the thermal treatment confirmed the validity of the diffusion model described in this paper. The carbon bulk concentration >100ppm is sufficient for the use of a steel material as ACS. The experimental results reported in this paper reveal, that with the MS-Probe the LOQ of less than one monolayer of iron oxide can be achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.