Abstract
Solid-state nuclear track detectors are widely used for integrated measurements of radon and its daughter products in the environment. However, these detectors have to be calibrated before use in field measurements. In the present work, LR-115 type II plastic track detectors have been calibrated in bare mode in a chamber containing a known radon concentration. The calibration factor measured for this mode was 125 tracks-cm^-2-d^-1 = 1 WL(3.12 X 10^-2 tracks-cm^-2 d^-1 = 1 Bq m^-3).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.