Abstract

The beta-gamma coincidence detector systems used for the measurement of the CTBT-relevant xenon isotopes (Xe-131m, Xe-133m, Xe-133 and Xe-135) in the International Monitoring System network and in the On-Site Inspection are reviewed. These detectors typically consist of a well-type or bore-through NaI crystal into which a measurement cell, serving also as a sample container, is inserted. This work describes the current calibration procedure for energy, resolution and efficiency, implementation challenges, availability and uncertainties of the specific nuclear decay data and the path forward to full calibration validation using GEANT4.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.