Abstract
This paper describes a measurement procedure developed by the Standards and Calibration Laboratory (SCL) for measuring the parameters of single phase 50 Ω//(50 μH + 5 Ω) V-network type line impedance stabilization networks (LISN) or artificial mains networks (AMN) specified in the international standard CISPR 16-1-2, which are the impedance at the equipment-under-test (EUT) port, the isolation between the power port and the receiver port, and the voltage division factor between the EUT port and the receiver port. The corresponding measurement models and evaluation of measurement uncertainties are presented. The characterization method of in-house developed adapters for BS1363 type plug and socket is also presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.